Shielding for Imaging Systems

Overview

Diagnostic X-ray system shielding is covered by NCRP-147 and AAPM TG-108. The methodology used for shielding imaging systems shares many similarities with linac vault shielding and NCRP-151.

Downloads

NCRP Report 147: Structural Shielding Design for Medical X-ray Imaging Facilities (External Link, requires AAPM membership)

AAPM TG-108: PET and PET/CT Shielding Requirements (External Link)

Typical Shielding Requirements

Room TypeTypical Shielding
High Energy Linear Accelerator Primary Barrier
  • 1.5-3m concrete
High Energy Linear Accelerator Secondary Barrier
  • 1-1.5m concrete
High Energy Linear Accelerator Door
  • With maze
    • 0.5 - 2cm inner lead layer
    • 2-4cm BPE
    • 1cm outer lead layer
  • No Maze
    • 5-9cm inner lead layer
    • 15-25cm BPE
    • 1cm outer lead layer
Ir-192 HDR Suite
  • ~50cm concrete
PET/CT Room
  • 1-2cm lead
  • 15-20cm concrete
CT Room
  • 1/16" to 1/8"Pb walls
  • 1/32" to 1/8" Pb ceiling and (possibly) floor
Radiographic Suite
  • Primary Barriers: 1/16"to 1/8"Pb
  • Secondary Barriers: 1/23"to 1/16" Pb

CT Shielding (NCRP 147 Methodology)

  • B is the maximum shielding transmission factor.
  • P are the shielding design goals.
    • Controlled areas: 0.1mGy/week, 5mGy/year
    • Uncontrolled areas: 0.02mGy/week, 1mGy/year.
  • is the distance from the bore center to the point of interest.
  • is the occupancy factor.
  • K1 is the air kerma per patient at a distance of 1 meter.
    • This may be determined using the CTDI100 method, the DLP method, or the isodose map method.
  • N is the number of patients per week.

Note that there is no use factor for a CT scanner. This is because CT rooms have no primary barriers because of the image detectors attenuation of the beam.

Key Point: NCRP-147 specifies shielding design goals in air kerma rather than dose equivalent or effective dose. This is because air kerma is readily measurable and the relationship between air kerma and effective dose depends on a variety of factors including orientation and posture of the individual.

NCRP-147 CT shielding thickness 24

Determining the thickness of shielding material required to reach a given transmission factor is more complicated than in MV shielding calculations because the attenuation coefficients of materials vary strongly with energy spectrum for kV photons. The below equation gives the minimum required barrier thickness. α, β, and γ are fitting parameters for broad beam x-ray source attenuation of a given shielding material.

Lead photon attenuation. Source: NCRP-147 figure A.2.
Concrete photon attenuation. Source: NCRP-147 figure A.3.

Methods for determining Air Kerma per patient (K1)

CTDI100 Method

This method uses measured values of
CTDI100 and an empirically derived κ value to determine K1.

  • P is the pitch
  • L is the scan length
  • κ is the scatter fraction per centimeter for head or body phantom
    • κhead = 9 x 10-5 cm-1
    • κbody = 3 x 10-4 cm-1

Dose Length Product (DLP) Method

Many modern CT scanners display dose-length-product (DLP) for their scans. This provides a simple approach to determining K1.

  • κ is the scatter fraction per centimeter for head or body phantom
    • κhead = 9 x 10-5 cm-1
    • κbody = 3 x 10-4 cm-1

Isodose Map Method

The simplest method for determining
K1 is to simply use a vendor supplied isodose map for the scanner.

If using this technique, care must be taken to correct for differences in acquisition technique used to create the map such as slice thickness, mAs, and kVp.

PET/CT Shielding

Shielding of a PET/CT scanner is the topic of TG-108. TG-108 focuses on F-18 based PET scanning because it is the most common radionuclide used in PET imaging. Additionally, because of F-18’s relatively long half life (110 minutes), shielding calculations based on F-18 are also adequate for other, shorter-lived radionuclides such as O-15 (T1/2 = 2 minutes), N-13 (T1/2 = 10 minutes) and C-11 (T1/2 = 20.4 minutes).

Key point: Typical PET/CT shielding requirements will be dominated by the 511keV photons that result from positron-electron annihilation from the PET radionuclide.

  • B is the maximum shielding transmission factor.
  • P the shielding design goal.
    • Controlled areas: 0.1mSv/week, 5mSv/year
    • Uncontrolled areas: 0.02mSv/week, 1mSv/year.
  • d is the distance from the center of the patient to the point of interest (m).
  • T is the occupancy factor.
  • NW is the number of patients per week.
  • A0 is administered activity (MBq).
  • tU and tI are the time in uptake room and time in imaging room respectively (hr).
  • FU is the uptake decay factor.
  • RtU and RtI are the dose reduction over uptake and imaging times.
    • Rt = 0.91 for 30 minutes
    • Rt = 0.83 for 60 minutes
    • Rt = 0.76 for 90 minutes

Navigation

Previous Page

Image Quality Metrics

Main Page

Table of Contents
General Medical Imaging Topics

Next Page

Table of Contents
Computed Tomography

Not a Premium Member?

Sign up today to get access to hundreds of ABR style practice questions.